Integrating Sphere Photometer Compact System for LED Lumen
Model: LPCE-3
LPCE-3 is a CCD Spectroradiometer Integrating Sphere Compact System for LED Testing. It is suitable for photometric, colorimetric and electrical measurement. It can measure LEDs, LED luminaires, Energy-saving lamps, Fluorescent lamps, HID lamps (high voltage sodium lamps and high voltage mercury lamps) and CCFL. The measured data meets the requirements of CIE 177, CIE84, CIE-13.3 , IES LM-79, Optical-Engineering-49-3-033602, IESNA LM-63-2, ANSI-C78.377 and IES LM 79-08 for the measurement of photometry and colorimetry.
System Configuration:
CCD Spectroradiometer (LMS-7000), Optical Fiber (CFO-1.5M), Digital Power Meter (LS2008R), DC Power Source (DC-S Series), AC Power Source (LSP-500VAS), Integrating Sphere (IS-1.5MA-CASE and IS-0.3M), Standard Light Source (SLS-50W and SLS-10W)
Measurement:
• Colorimetric: Chromaticity coordinates, CCT, Color Ratio, Peak Wavelength, Half Bandwidth, Dominant Wavelength, Color Purity, CRI (Ra, R1 to R15), Spectrum Test, TM30 (Rf, Rg), CQS
• Photometric: Luminous Flux, Luminous Efficiency, Radiant Power, EEI, PAR, PPF
• Electrical: Voltage, Current, Power, Power Factor, Harmonic (Optional)
Specification:
• Spectral Range Wavelength: 350nm~950nm(LMS-700VIS) (200~1050nm is optional)
• Spectral Wavelength Accuracy: ±0.5nm Wavelength Reproducibility: ±0.5nm
• Sample Scanning Steps: ±0.1nm
• Accuracy of Chromaticity Coordinate (Δx, Δy): ±0.003
• Correlated Color Temperature CCT: 1,500K~25,000K. CCT Accuracy: ±0.5%
• Color Rendering Index Range: 0~100.0, Accuracy: ± (0.3%rd±0.3)
• Photometric linear: ±0.5%
• Stray light: <0.015% (600nm) and <0.03% (435nm)
• Integrtion Time: 0.1mS~20S