1. |

LISUN

Home> Product> Environmental Test Chamber

    ASTM Sulfur Dioxide Test Chamber Machine

    Model: SQ-010

    Get a Quote

    Your email address will not be published. Required fields are marked *

    1. Description

    This Sulfur Dioxide Test Chamber Machine is designed according to GB/T2423.19-1981, GB/T2423.33-2005, GB/T5135.3-2003, ISO6988 (replaced DIN50018: 1988-06), DIN50900, DIN53210 and EC, MIL, ASTM etc to do sulfur dioxide gas corrosion test.

    Model Work Room Size (mm) Exterior Size (mm) Voltage Power (KW)
    SQ-150 450X600X400 800x1070x1050 AC220 2.5
    SQ-250 600x900x500 870x1400x1150 AC220 3
    SQ-750 750x1100x500 1030x1680x1260 AC220 4
    SQ-010 850x1300x600 1160x2000x1390 AC380 6
    SQ-016 850x1600x600 1280x2320x1500 AC380 7.5
    SQ-020 900x2000x600 1220x2530x1410 AC380 9


    Specifications:
    • Temperature range: RT+5~55℃ (Can be customized according to test demand)
    • Temperature fluctuation: ±0.5℃
    • Temperature uniformity: ±2℃
    • Relative humidity: 75~98%RH
    • Sulfur dioxide concentration range (Please mark this on the PO): 0~5000ppm (0~1000pphm adjustable)
    • Quantitative control (Please mark this on the PO): 0~10L/hour (Adjustable)
    • Environmental Requirements:
    1)Power Source: AC380V 50/60Hz
    2)The distance between cabinet and surface of wall should be more than 600mm
    3)The left of machine should be close to wall, and the outside of test room should be outdoor to emissions of sulfur dioxide gas
    4)This machine should put a room separate, and this room should be easy to ventilation with outdoor
    • Machine structure and characteristics of the material:
    1)Box with glass fiber reinforced epoxy molding touch, abestos jacket stickers, endure high temperature and corrosion resistance, light weight, high strength, smooth and easy to clean wall cabinets
    2)Cover with epoxy fiberglass roof made of 100 ~ 120 ℃, in order to avoid condensation inside directly onto the specimen. Open the lid attached pneumatic devices
    3)Heater: Nickel-chromium alloy heating wire is heated from the bottom, forming a jacket heating to keep the temperature uniformity is good
    4)Humidification method: Humidifying steam generator
    5)Sulfur dioxide input method: Cylinders input (Manual)
    6)Exhaust gas processing (forced discharge function): Install sulfur dioxide gas process device, multi-stage process by sodium bicarbonate, to pass emissions (purification devices external)

    1. Document