    1. Description

    Basic Introduction
    DSCD-900 is a highly performance loaded spectrophotometer with test caliber upward. The instrument adopts the geometric conditions of D/8 lighting and observation of CIE recommendations, integration of the full band LED composite light source, double beam optical system, ETC every-time calibration technology, spectral reflectance standard deviation within 0.08, maximum of colorimetric values ΔE*ab is 0.03, to provide all the necessary color formula. It could be widely used for coating, printing, textile, meet plastic, food, medicine, automobile field to meet customers’ measurement requirement.


    Technical Features
    • Bearing type test
    • D\8 Structure
    • Can Measure SCI and SCE at the same time
    • Download
    • Color Quality Control software


    Technical Data

    Type DSCD-900
    Illumination d/8 (Diffused illumination, 8°viewing) (conform to CIE No.15、ISO 7724/1、ASTM E1164、DIN 5033 Teil7、JIS Z8722 Condition c standards)
    Size of integrating sphere Φ40mm, Alvan diffused reflection surface coating
    Illumination Light source CLEDs (entire wavelength balanced LED light source)
    Sensor dual optical path sensor array
    Wavelength range 400-700nm
    Wavelength interval 10nm
    Half spectral width 5nm
    Reflectivity range 0-200%
    Resolution 0.01%
    Observer angle 2°/10°
    Measurement Light Source A, C, D50, D55, D65, D75, F1, F2, F3, F4, F5, F6, F7, F8, F9, F10, F11, F12, DLF, TL83, TL84, NBF, U30, CWF
    Data being displayed Reflectance graph/value, chromatic value, color difference values, color assessment results, color tendency, display measurement area, history color value simulation, manual input standard sample, generate measurement reports
    Measurement time About 1 second
    Measurement time interval 0.5 second
    Measuring caliber Φ11mm
    Color space CIE-L*a*b, L*C*h, L*u*v, XYZ, Yxy, reflectivity, Hunterlab, Munsell MI, CMYK, RGB, HSB
    Color difference formulas ΔE*ab, ΔE*CH, ΔE*uv, ΔE*cmc (2:1), ΔE*cmc (1:1), ΔE*94, ΔE*00, ΔEab (Hunter), color classification
    Other colorimetric indices WI (ASTM E313-00, ASTM E313-73, CIE/ISO, AATCC, Hunter, Taube Berger Stensby), YI (ASTM D1925, ASTM E313-00, ASTM E313-73), Tint (ASTM E313, CIE,Ganz), metamerism index Milm, adhesive/changing color fastness, ISO luminance, 8 gloss, A density, T density, E density, M density
    Repeatability light splitting reflectivity: standard deviation within 0.08%
    color values: ΔE*ab <=0.015 (After calibration, standard deviation of 30 measurements on test white board, 5 second intervals), Maximum: 0.03
    Battery capacity rechargeable, 20000 continuous tests, 7.4V/6000mAh
    Interface USB
    Data storage 40000 test results
    Light source longevity 5 years, 1.5 million tests
    Inter-instrument agreement ΔE*ab within 0.2 (BCRA color charts II, average of the 12 charts)
    Size 350*300*200mm (L*W*H)
    Weight 4.5kg
    Display 5 inches True color screen
    Work temperature range 0~45℃, relative humidity 80% or below (at 35°C), no condensation
    Storage temperature range -25℃ to 55℃, relative humidity 80% or below (at 35°C), no condensation
    Standard accessories AC power line, operating manual, color QC software, driving software, electric operating manual, USB cable, white/black calibration tile, verification certification
    Optional accessories powder molding device, cuvette, shading cover